abruptly, thus giving rise to doubt as to which limits apply, the
manufacturer may specify any value which is within 10% of the actual
measured value.
The value for power found by measurement during emission tests other
than the original manufacturer's conformity test, measured according to the
terms of this clause, shall not be less than 90% nor greater than 110% of
the value for power specified by the manufacturer….. In the event that the
measured value is outside of this tolerance band about the specified value,
the measured power shall be used to establish limits.
If the 'specified' power is unknown, it will be measured by the analyzer and
this value can be entered retrospectively at the end of the tests and the
limits re-computed accordingly.
IEC61000-3-2 2001 Fundamental Current
The fundamental current is used to calculate some Class C limits. This field
is inactive when Class A, B or D is selected.
The fundamental input current is defined to be the maximum of the
measured values of 1st harmonic current from the DFT time windows over
the entire duration of the test.
If the fundamental current is unknown, it will be measured by the analyzer
and this value can be entered retrospectively at the end of the tests and the
limits re-computed accordingly.
IEC61000-3-2 2001 Power Factor
The power factor is used to calculate some Class C limits. This field is
inactive when Class A, B or D is selected.