measures a mean temperature of 250°C or that available temperature
closest to this value.
Microwave ovens
The microwave oven or the microwave function of a combination oven
shall be tested at the lowest, medium and a third stage that is the highest
adjustable power less than or equal to 90% of the maximum power. Load
the oven with a glass bowl containing 1000 ± 50g of water.
Test conditions for lighting equipment
The following test conditions shall apply to equipment with a primary
function of generating and/or regulating and/or distributing optical
radiation by means of incandescent or discharge lamps or LEDs.
Such equipment shall be tested with a lamp of that power for which the
equipment is rated. If lighting equipment includes more than one lamp, all
lamps shall be in use.
Pst and Plt evaluations are required only for lighting equipment that is
likely to produce flicker; for example: disco lighting and automatically
regulated equipment.
No limits shall apply to lamps.
Incandescent lamp luminaires with ratings less than or equal to 1000W and
discharge lamp luminaires with ratings less than or equal to 600W, are
deemed to comply with the dmax limits in this standard and are not
required to be tested. Luminaires with higher ratings, which cannot comply
with this part of IEC61000, shall be subject to conditional connection in
accordance with IEC61000-3-11.
Ballasts are deemed to be part of luminaires and are not required to be
tested.