Page 58
VPN 98-025/7
Voltech Instruments
click on the Flicker Meter button to display:
Normally, for full compliance testing, you would set the Measurement
Method to Voltage, which would require you to insert a standard
impedance network in the power line to the product under test.
However, assuming that this time you want to use the measurement
rig shown above for pre-compliance testing, click on the Current
button in the Measurement Method box; the Resistance and Reactance
default to the normal values specified in the IEC61000-3-3 Standard.
For the 2001 edition of the standard,
The dc limit is 3.3%
The dt limit is 500mS (the time for which the voltage characteristic
exceeds a voltage change of 3.3%)
The dmax limit is:
a) 4% without additional conditions
b) 6% for equipment that is:
- Switched manually
OR
- Switched automatically > twice per day