The maximum permissible harmonic currents expressed as a percentage of
the input current at the fundamental frequency are:
3rd 86 %
5th 61 %
Note that for this test, the fundamental input current must be known and
entered in to the test detail dialogue since this value is used to calculate the
limits. The analyzer automatically measures fundamental input current in
accordance with the required measurement methods and displays the result.
The measured fundamental input current can then be entered
retrospectively and the limits re-computed.
IEC61000-3-2 2001 Waveform Check
For a 'waveform check' the analyzer performs a waveform datalog and
displays one cycle of the voltage and current.
This waveform can then be used to check if Class C 'by waveform' limits
can be used by pressing the space bar. The markers show the angular
position of current 'flow start', 'last peak' and 'flow stop'.