Other Data Plot
These views show data other than harmonics as they vary during the
duration of the test.
Watts and Amps
Shows the input watts and rms amps as measured by the analyzer. Input
Watts is used to calculate limits for Class D tests and the user specified
watts are also shown for comparison. The maximum measured watts should
be within 10% of the user specified watts.
Click anywhere on this window to move the red marker and view details
for the associated reading at the top of the main window.
Amps and Power Factor
This view shows the fundamental input current (amps) and power factor as
they varied during the test together with their user specified levels. See the
reference information for definitions of fundamental current and power
factor.
Fundamental current and power factor are used to calculate limits for some
Class C tests. The maximum measured fundamental current and power