IEC61000-3 S
OFTWARE FOR
W
INDOWS
U
SER
M
ANUAL
R
EFERENCE
I
NFORMATION
Voltech Instruments
VPN 98-025/7
Page 85
Lamps
Lamps shall be aged for at least 100h at rated voltage. They shall be
operated for at least 15 minutes before a series of measurements is made.
During ageing and measurement lamps shall be installed as in normal use.
Note - Some lamp types may require a stabilising period exceeding 15
minutes. Corresponding information given in the relevant lamp
specification must be observed.
Luminaires
The luminaire is measured as manufactured. It shall be tested with
reference lamps, or with lamps having electrical characteristics close to
their nominal values. In case of doubt measurements are made with
reference lamps. When the luminaire incorporates more than one lamp, all
lamps are connected and operated during the test. When the luminaire is
assigned for use with more than one type of lamp, measurements shall be
made with all the types and the luminaire shall comply each time. In each
case where the luminaire is equipped with a glow starter, a starter in
accordance with IEC Publication 155 shall be used.
Incandescent lamp luminaires that do not incorporate an electronic
transformer or a dimming device are deemed to fulfil the harmonic current
requirements and need not be tested.
If separate tests with reference lamps have proved that ballasts for
fluorescent or other discharge lamps or step-down converters for tungsten
halogen or other filament lamps comply with the requirements the
luminaire is deemed to comply with these requirements and need not be
checked. Where these components have not been approved separately or do
not comply, the luminaire shall be tested and shall comply.