134
IEC 61000-4-4
±1 kV for
input/output lines
±1 kV for
input/output lines
environment.
±1 kV differential
mode
±1 kV differential
mode
Surge
Mains power quality should be that of
a typical commercial or hospital
environment.
IEC 61000-4-5
±2 kV common
mode
±2 kV common
mode
<5 % UT
<5 % UT
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
(>95 % dip in U
T
)
(>95 % dip in U
T
)
for 0,5 cycle
for 0,5 cycle
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
Digitizer requires continued operation
during power mains interruptions, it is
recommended that the Digitizer be
powered from an uninterruptible power
supply or a battery
40 % U
40 % U
T
T
)
)
(60 % dip in U
(60 % dip in U
T
T
for 5 cycles
for 5 cycles
70 % U
70 % U
T
T
)
)
(30 % dip in U
(30 % dip in U
T
T
for 25 cycles
for 25 cycles
<5 % U
<5 % U
T
T
(>95 % dip in U
T
)
for 5 sec
(>95 % dip in U
T
)
for 5 sec
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
Power frequency
(50/60 Hz)
3 A/m
3 A/m
magnetic field
IEC 61000-4-8
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
VIDAR Advantage Series Installation and Operation Guide
Summary of Contents for CAD PRO Advantage
Page 5: ......
Page 12: ...7 DIAGNOSTICPRO Advantage and DOSIMETRYPRO Advantage Installation Guide ...
Page 30: ...25 CADPRO Advantage Installation Guide VIDAR CADPRO Advantage Installation Guide ...
Page 47: ...42 VIDAR Advantage Series Installation and Operation Guide ...
Page 48: ...43 Interfacing Guide VIDAR Advantage Series Installation and Operation Guide ...
Page 49: ...44 VIDAR Advantage Series Installation and Operation Guide ...
Page 87: ...82 VIDAR Advantage Series Installation and Operation Guide ...
Page 88: ...83 Advantage Series Operating Guide VIDAR Advantage Series Installation and Operation Guide ...
Page 97: ...92 VIDAR Advantage Series Installation and Operation Guide ...
Page 102: ...97 Advantage Series Other Information VIDAR Advantage Series Installation and Operation Guide ...
Page 143: ...138 VIDAR Advantage Series Installation and Operation Guide ...