Electromagnetic guidance
Caution: Medical electrical equipment.
EMC (Electro Magnetic Compatibility) must be considered before any medical electrical equipment
is installed or put into service. Follow the information in the accompanying documentation when
installing and operating the Digitizer.
Caution: Portable or mobile RF communication equipment can effect Medical Electrical equipment.
Caution: Using the Digitizer adjacent to or stacked with other equipment may cause interference
between the equipment. Before utilizing stacked or adjacent equipment, verify proper functionality
of all equipment in the actual configuration in which it will operate.
Caution: Connecting the Digitizer to equipment that is not rated CISPR 11 class A or class B may
alter the electromagnetic characteristics.
Caution: In a domestic environment this product may cause radio interference in which case the
user may be required to take adequate measures.
Guidance and manufacturer’s declaration –
electromagnetic equipment
Guidance and manufacturer’s declaration – electromagnetic immunity
Table 202
T
for use in the e
e customer or end user should assure that
they are used in such an environment.
he D
IAGNOSTIC
P
RO
Advantage, D
OSIMETRY
P
RO
Advantage and C
AD
P
RO
Advantage Digitizers are intended
lectromagnetic environment specified below. Th
Emi
ssions test
Compliance
Electromagnetic
ent - guidance
environm
RF Emissions
CISPR11
Group 1
The Digitizers us
here
and are not likely
nic equipm
e RF energy only for their internal
fore, their RF emissions are very low
to cause any interference in nearby
ent.
functions. T
electro
RF Emissions, CISPR11
Class A
H
nic Emissions
armo
IEC 61000-3-2
Class A
Voltage Fluctuations/
flicker emissions
IEC 61000-3-3
Complies
The Digitizers are
ents
other than domestic and those directly connected to the
public low voltage power supply network that supplies
buildings used for domestic purposes.
suitable for use in all establishm
The D
IAGNOSTIC
P
RO
Advantage, D
OSIMETRY
P
RO
Advantage and C
AD
P
RO
Advantage Digitizers are
intended for use in the electromagnetic environment specified below. The customer or end user of the
Digitizer should assure that it is used in such an environment.
Table 201
IEC 60601 test
level
Electromagnetic environment -
guidance
Immunity test
Compliance level
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
ElectroStatic Discharge
(ESD)
+ 6 kV contact
+ 6 kV contact
IEC 61000-4-2
+ 8 kV air
+ 8 kV air
Electrical fast transient/
burst
±2 kV for power
supply lines
±2 kV for power
supply lines
Mains power quality should be that of
a typical commercial or hospital
VIDAR Advantage Series Installation and Operation Guide
Summary of Contents for CAD PRO Advantage
Page 5: ......
Page 12: ...7 DIAGNOSTICPRO Advantage and DOSIMETRYPRO Advantage Installation Guide ...
Page 30: ...25 CADPRO Advantage Installation Guide VIDAR CADPRO Advantage Installation Guide ...
Page 47: ...42 VIDAR Advantage Series Installation and Operation Guide ...
Page 48: ...43 Interfacing Guide VIDAR Advantage Series Installation and Operation Guide ...
Page 49: ...44 VIDAR Advantage Series Installation and Operation Guide ...
Page 87: ...82 VIDAR Advantage Series Installation and Operation Guide ...
Page 88: ...83 Advantage Series Operating Guide VIDAR Advantage Series Installation and Operation Guide ...
Page 97: ...92 VIDAR Advantage Series Installation and Operation Guide ...
Page 102: ...97 Advantage Series Other Information VIDAR Advantage Series Installation and Operation Guide ...
Page 143: ...138 VIDAR Advantage Series Installation and Operation Guide ...