108
Top LED: feeder status
LED state
Digitizer condition Corrective action
Solid green
Ready
None. The feeder is ready to operate.
Off
ERROR: no feeder
attached
1. Turn off digitizer.
2. Remove feeder.
3. Inspect electrical connector and screws on
bottom of feeder.
Fast flashing green
ERROR: unknown
feeder attached
Check feeder:
1. Turn off digitizer.
2. Remove feeder.
3. Inspect electrical connector and screws on
bottom of feeder.
Slow flashing green
BUSY: film is
staging
None. This is normal operation.
Fast flashing green
ERROR: film staging
problem
Eject film by clicking the eject button in the
scanning application.