110
Bottom LED: power status
LED state
Digitizer condition Corrective action
Solid green
Ready
None.
Fast flashing green
ERROR: DSP not
communicating
Get qualified technical help (see “In case of
difficulty,” later in this chapter).
Fast flashing green
ERROR: unknown
ballast
Get qualified technical help (see “In case of
difficulty,” later in this chapter).
Off
ERROR
1. Make certain power is supplied to the digitizer,
and that the digitizer power switch is on.
2. If problem persists, get qualified technical help
(see “In case of difficulty,” later in this chapter).