2.8 Pipe installation
Microwave can propagate only inside of metal pipe. Following features are for the application for pipe
installation.
・
It can use existing metal pipe without replacement.
・
It will not be affected by any obstacle reflection disturbance outside of the pipe.
・
The deterioration of the signal is much smaller than free space propagation. Thus even low
dielectric constant liquid may have longer measurable range.
・
Pipe can be used as stilling well against turbulence.
NOTE:
・
Antenna should be selected 4” or 2” cone antenna type.
・
Pipe material should be metal.
・
Pipe length should cover all measuring range.
・
The gap between antenna outer diameter and pipe inner diameter should be less than 2 mm.
・
Air vent hole (approx. 5mm-10mm) must be positioned above maximum liquid level.
・
Low dielectric constant liquid may require to attach bottom reflection plate as Fig. 2.2.12.
・
In case target liquid is non-homogeneous characteristic or has multiple layer, some of slits may be
required for mixing each in and out side liquid. Those slits should be less than 5mm and positioned
right-angled against “
E
”.
・
In case target liquid has high viscosity, periodical cleaning maintenance may be required.
・
In case the pipe will have strong stress, pipe should be fixed to wall or protection arm.
・
Inside pipe, between flange connections or welding requires smooth finish.
100 %
0 %
Stirring hole
Air vent hole
Reflector
View A-A
A
A
Air vent hole
Fig. 2.8.1 Pipe installation
Summary of Contents for KRG-10
Page 1: ...Document No KF13 001K Microwave Level Gauge RG 10 User s Guide...
Page 9: ...8 Blank...
Page 13: ...12 Blank...
Page 51: ...3 8 Blank...
Page 55: ...4 4 Blank...
Page 171: ...9 2 Blank...
Page 190: ...10 19 M20 1 5 Fig 10 7 6 Cable gland Only Non Ex type...
Page 193: ...11 2 Blank...
Page 197: ...12 4 Blank...
Page 199: ...Blank...