6-81
(1) Executing Parameter Initialization(*It is supported by firmware version between 1.4.1
and less than 1.5.0)
To initialize device parameters, select "Parameter Initialization" ( ), and then click the
"Execute" button. The following window will be displayed, so then click "OK" to initialize the
device parameters.
Fig. 6.10.2 Parameter initialization confirmation dialog box
The parameters that are initialized by config software
Parameters except the noise table, fixed disturbance echo, and HART bus address.
(2) Executing Parameter Backup(*It is supported by firmware version 1.5.0 or later)
To backup device parameters, select "Parameter Backup" ( ), and then click the "Execute"
button. The following window will be displayed, so then click "OK", the setting parameters are
stored in the inside of the device. When parameter backup is executed, you have to be very
careful because factory-configured parameters are overwritten and removed.
Fig. 6.10.3 Parameter backup confirmation dialog box
Please don't turn off power for five seconds after executing parameter backup because it leads
to device failure.
The parameters that are initialized by config software
Parameters except the noise table, fixed disturbance echo, and HART bus address.
(3) Executing Parameter Recovery(*It is supported by firmware version 1.5.0 or later)
To recovery device parameters, select "Parameter Recovery" ( ), and then click the
"Execute" button. The following window will be displayed, so then click "OK", the parameters
stored in the inside of the device are loaded. When parameter backup is never executed,
factory-configured parameters are loaded.
Summary of Contents for KRG-10
Page 1: ...Document No KF13 001K Microwave Level Gauge RG 10 User s Guide...
Page 9: ...8 Blank...
Page 13: ...12 Blank...
Page 51: ...3 8 Blank...
Page 55: ...4 4 Blank...
Page 171: ...9 2 Blank...
Page 190: ...10 19 M20 1 5 Fig 10 7 6 Cable gland Only Non Ex type...
Page 193: ...11 2 Blank...
Page 197: ...12 4 Blank...
Page 199: ...Blank...