6-46
(4) 20mA calibration
To perform a 20mA calibration on the device, click the "20mA Calibration" button
②
. The
following window will be displayed, so then click "OK" to output a fixed current of 20mA.
Fig. 6.7.6 20mA calibration confirmation dialog box
Fig. 6.7.7 20mA output
Then the following window will be displayed, so follow the window display to measure the output
current. After clicking "OK" once measurement has finished, input the calibration value (present
current value) and click the "Write" button to write the calibration value to the device.
Fig. 6.7.8 Measurement / calibration value setting instructions dialog box
Fig. 6.7.9 Input 20mA calibration value
The 20mA calibration is now finished.
Analog data
- Fixed Current Output Fixed current that is output
(ON
(3.6mA
to
22.0mA)/OFF
(0mA))
- 4mA Calibration
4mA calibration value of analog current output
(3.5mA
to
4.5mA)
- 20mA Calibration
20mA calibration value of analog current output
(15.0mA
to
25.0mA)