5.6.2
Guiding Capillaries and Tubing Through the System
Flow connections between the modules of the Vanquish system are
guided through either the tubing chase in the devices or the guide holes
or capillary clips of the devices.
Tubing Chase, Tubing Guide, Tubing Bracket
To guide certain tubes and lines (solvent tubing, wash liquid tubing,
detector waste line) from the top module to the bottom module in the
Vanquish system stack, the stackable modules have a tubing chase on
the inside right.
The tubing chase provides four tubing guides. Each guide can hold up to
three tubes or lines.
In each module, push the tube (or line) into the appropriate guide (see
picture).
Figure 15: Tubing chase with tubing guides (left: view from inside, right:
view from top)
No.
Use for
1
Solvent tubing (up to three solvent lines)
2
Solvent tubing (up to three solvent lines)
3
Wash liquid tubing (seal wash, autosampler needle wash)
4
Detector waste line
Summary of Contents for Vanquish VH-D10
Page 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Page 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...