Parameters
Description
Analog out board
parameters
For configuring the analog out connections, parameters
are available for Analog Out port 1 or 2, respectively.
Analog1_Offset
Analog2_Offset
Adjusts the zero position of the analog output when it is
plotted. The value entered is a percentage of the full-scale
analog output. The offset allows the evaluation device to
plot the signal if it becomes negative.
Note that the output voltage of the analog outputs cannot
reach a minimum of exactly 0 V. Therefore, we
recommend to set an offset of at least 5%, even if only
positive analog output signals are expected. This will
enable you to also measure very small absorptions down
to 0 AU precisely.
Analog1_Range
Analog2_Range
Sets the voltage range of the analog output signal (1 V or
10 V). The range to use depends on the evaluation device
to which the output is connected.
Analog1_Resolution
Analog2_Resolution
Sets the resolution of the signal. The resolution to use
depends on the maximum absorption value expected for
the application and relates to a full-scale output when
AnalogX_Offset
is 0%.
Summary of Contents for Vanquish VH-D10
Page 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Page 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...