Follow these steps
1. When removing the waste line from the tubing guides, be careful
not to pull on other tubing in the guides.
2. Disconnect the waste line from the flow cell outlet.
3. Remove the waste line from the top recess of the partition panel:
♦
If the waste line is installed in a rotating plug
Turn the plug in the top recess to the front and remove the
waste line from it.
♦
If the waste line is installed in a plug with slit
Pull the waste line out of the slit.
4. Remove the waste line from the detector waste port in the system
base and from waste.
5. Unpack the replacement waste line.
6. Set up the waste line. Follow the instructions in
section 5.6.4,
page 67