Specifications
Vanquish Diode Array Detectors (VH-D10) Operating Manual
Page 177
9.1.2
LightPipe Flow Cells
The flow cell performances are specified as follows:
Type
10 mm Fused-Silica LightPipe Flow
Cell Specification
60 mm Fused-Silica LightPipe Flow
Cell Specification
Design
Light pipe
Light pipe
Path length
10 mm
60 mm
Illuminated volume
2 µL
13 µL
Dispersion volume
0.8 µL
4.0 µL
Pressure limit
6 MPa
6 MPa
Max. eluate
temperature limit
50 °C
50 °C
Biocompatibility
yes
yes
Materials in the
analytical flow path
Fused silica, PEEK, perfluoro-
elastomer, titanium
Fused silica, PEEK, perfluoro-
elastomer, titanium
Summary of Contents for Vanquish VH-D10
Page 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Page 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...