6
•
Operation
Vanquish Diode Array Detectors (VH-D10) Operating Manual
Page 85
MUTE ALARM
If beeping alerts you that the detector firmware detected a problem, for
example leakage, beeping continues until you turn off the beep. Pressing
this button turns off the beep for the current alarm. As a standard,
beeping starts again after 10 minutes if the problem is still present or if
the firmware detected another problem.
UV
The
UV
button allows you to turn the UV lamp on and off directly from
the detector. To turn on the UV lamp, press the
UV
button.
The LED next to the button indicates the lamp status:
LED
Description
Off (dark)
The UV lamp is turned off.
Green, flashing
The UV lamp ignites.
Green
The UV lamp is turned on.
When the detector is connected in the Chromeleon software
The button functionality is as follows when the detector is connected in
the Chromeleon software:
•
No injection (sample) or sequence or baseline monitoring is running:
All functions are available from the keypad.
•
An injection (sample) or sequence or baseline monitoring is running:
The
Mute Alarm
function remains available from the keypad,
allowing you to turn off the beep for the current alarm.
Summary of Contents for Vanquish VH-D10
Page 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Page 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...