Shutting Down the Detector
If the detector will not be operated for some time, follow the
instructions on shutting down the detector in this section.
TIP
The Chromeleon software provides procedures for automatically
preparing the chromatography system for shutdown. The procedures
include, for example, operations for reducing the flow rate, reducing the
temperature in temperature-controlled devices, and turning off the
detector lamps.
For information about
Smart Shutdown
and
Smart Standby
, refer to the
Chromeleon Help
.
6.8.1
Short-Term Shutdown (Interruption of Operation)
To interrupt operation of the detector for a short period (short-term
shutdown), for example, overnight, observe these guidelines for the
Vanquish system modules, as required by your system arrangement:
•
For your Vanquish detector, note the following:
♦
Charged aerosol detector:
Check that sufficient gas is available to continue gas flowing
through the detector. This is to prevent any build-up of residue
from solvents or analytes. Gas must be flowing when pump flow
is delivered to the detector.
♦
UV/VIS detector:
The lamp(s) in the detector can remain turned on. The shutter
can be moved to a closed position for protection of the flow cell.
♦
Fluorescence detector:
Turn off temperature control for the flow cell.
•
Apply a flow of 0.05 mL/min and have the pump deliver an
appropriate solvent.
Check the lower pressure limit for the pump and adapt the value if
necessary. If the pressure falls below the lower limit, the pump stops
the flow.
•
Set the injection valve in the autosampler to the Inject position.
•
Make sure that the temperature of the column does not exceed
40 °C.
Summary of Contents for Vanquish VH-D10
Page 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Page 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...