7
•
Maintenance and Service
Page 138
Vanquish Diode Array Detectors (VH-D10) Operating Manual
Back-flushing
1. Disconnect the back-flush waste line from the back-flush capillary.
2. Connect the Viper connector to the T piece.
3. Connect the back-flush waste line to the flow cell inlet.
4. Route the free end of the back-flush waste line to waste.
5. Connect the Viper connector of the back-flush kit to the flow cell
outlet.
Figure 34: Setup for back-flushing
No.
Description
1
To the outlet of the pump
2
Back-flush waste line
3
Waste container
6. Turn on the pump flow at the flow rate as determined in
section
Determining the required pump flow
7. Write down the pressure at the beginning of the flushing.
8. Flush the flow cell for 2 minutes with the back-flushing agent.
9. Stop the pump flow.
10. Forward flush the flow cell (see
section
Forward flushing below
).
Summary of Contents for Vanquish VH-D10
Page 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Page 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...