Troubleshooting
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
Page 161
Message and Code
Description and Remedial Action
Code 8111 - Code 8115 Calibration of
xx failed.
With x = Name of the component for which calibration failed
Cause details
Raman calibration failed.
Remedial actions
•
Observe the conditions for wavelength calibration (see
) and wavelength validation
(see
Performing a Wavelength Validation (
•
Verify that the flow cell is installed correctly and that the screws are
tightened finger-tight.
•
Clean the flow cell (see
) if
necessary.
•
Make sure that the flow cell is properly equilibrated and no air
bubbles are present. Make sure that you use degassed solvents.
If the procedure is repeatedly interrupted, contact Technical Support.
Code 8116 and 8118 to 8121
Initialization failed
Cause details
•
The detector could not initialize.
•
The flow cell may be contaminated.
Remedial actions
•
Verify that a flow cell is installed correctly.
•
Verify that the eluent composition in the flow cell does not change
and no air bubbles are present. Make sure that you use degassed
solvents.
•
Clean the flow cell (see
) if
necessary.
Summary of Contents for Vanquish VC-D50
Page 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...