Operating Manual
5. Stop the pump flow.
6. Calculate the difference between the two measured pressure values:
p5 = p4 – p1.
p5: Pressure drop at the flow cell
p4: Pressure drop at the flow cell including the column, detector
inlet capillary and waste line
p1: Pressure drop at the column, detector inlet capillary and the
waste line
5.7.5
Calculating the Pressure at the Inlet Port of the Flow Cell
This procedure describes how to determine the pressure at the flow cell
inlet port (p6) for the intended configuration.
Figure 31: Calculating the pressure at the inlet port of the flow cell in
your used system configuration (here with second detector)
No.
Description
No.
Description
1
Detector
5
Waste container
2
Flow cell
6
Outlet port of the second
detector
2a
Outlet port of the flow cell
7
Inlet port of the second
detector
2b
Inlet port of the flow cell
8
Second detector
3
Detector inlet capillary
9
Waste line
4
Column
10
Transfer capillary
Summary of Contents for Vanquish VC-D50
Page 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...