5
•
Installation
Page 68
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
Figure 20: Capillary installed in the plug with slit
Installing the rotating plug
1. On the detector partition panel, push the rotating plug in the
required recess of the partition panel (if not present yet).
2. To open the plug in order to route the capillary through the plug,
turn the rotating plug toward the front.
3. To secure the capillary, turn the rotating plug toward the partition
panel to close the opening of the rotating plug.
Figure 21: Using the rotating plug
5.6.4
Connecting Fittings, Capillaries, and Tubing
The inlet and outlet ports of the flow cells support the following capillary
fittings:
•
Viper™
•
nanoViper™
•
Standard 1/16" HPLC fittings using ferrules (PEEK, SST) or finger-tight
fittings (PEEK)
This section provides information about how to connect and handle
capillaries, fittings, and tubing.
Summary of Contents for Vanquish VC-D50
Page 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...