6
•
Operation
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
Page 117
6.9.10 Baseline Behavior
If wavelengths are switched or the sensitivity, filter wheel, or PMT
setting are changed during the method, the chromatogram may show
baseline jumps because of the background fluorescence at the new
wavelength setting. Three baseline behavior mode settings define how
the baseline behaves in these cases:
•
Zero
: sets the baseline to zero
•
Append
: appends the baseline to the previous signal
•
Free
: lets the baseline "jump" to the current absolute value
TIP
You can perform
Autozero
command (automatic null balancing) any
time, that is, the intensity measured at the time of the autozero is
subtracted from any intensities measured thereafter. Always select a
time for null balancing at which no sample flows through the flow cell.
The values are stored and can be reset using the
ClearAutozero
command.
Summary of Contents for Vanquish VC-D50
Page 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...