Installation
Page 70
Fluorescence Detectors (VC-D50, VC-D51, VF-D50, VF-D51)
Operating Manual
Figure 22: Viper fitting with knurl
No.
Description
1
Knurl
2
Capillary
3
Slot
1. Insert the Viper capillary into the connection port.
2. Tighten the connection by the knurl.
TIP
Note the slot in the knurl. For narrow connections, you can easily
remove the knurls from neighboring capillaries through this slot and
attach them again later.
3. Check whether the connection leaks. If leakage exists, follow the
steps further down.
Resolving Leakage of Viper Fittings with Knurls
1. Tighten the connection a little more.
2. If leakage continues, remove the capillary.
3. Clean the capillary ends carefully by using a lint-free tissue wetted
with isopropanol.
4. Reinstall the capillary.
5. If the connection continues to leak, install a new Viper capillary.
Summary of Contents for Vanquish VC-D50
Page 4: ...Contacting Us Page 4 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 10: ...Contents Page 10 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 30: ...2 Safety Page 30 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 42: ...3 Device Overview Page 42 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 88: ...5 Installation Page 88 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 122: ...6 Operation Page 122 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 170: ...9 Specifications Page 170 Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual...
Page 187: ...Index Fluorescence Detectors VC D50 VC D51 VF D50 VF D51 Operating Manual Page 187...