Chapter 1 Introduction
SPECIFICATIONS
Preset ranges
0-1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000, 5000,
10000 ppm
0-1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000, 5000,
10000 mg/m
3
Custom ranges
0-1 to 10000 ppm
0-1 to 10000 mg/m
3
Zero noise
0.02 ppm RMS (30 second time setting)
Lower detectable limit
0.04 ppm
Zero drift (24 hour)
< 0.1 ppm
Span drift (24 hour)
±
1% fullscale
Response time
60 seconds (30 second time setting)
Precision
±
0.1 ppm
Linearity
±
1% fullscale
≤
1000 ppm
±
2.5% fullscale > 1000 ppm
Sample flow rate
0.5-2 liters/min
Operating temperature
20 - 30
°
C (may be safely operated over the range of 0 -
45
°
C)*
Power requirements
105-125 VAC, 60 Hz
220-240 VAC, 50 Hz
100
Watts
Physical dimensions
16.75" (W) X 8.62" (H) X 23" (D)
Weight 45
lbs.
Outputs CO
selectable
voltage
4-20 mA, RS-232, RS-485
* In non-condensing environments
1-3
Summary of Contents for 48C
Page 4: ......
Page 12: ......
Page 14: ......
Page 18: ......
Page 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Page 112: ......
Page 116: ......
Page 120: ......
Page 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Page 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Page 138: ......
Page 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Page 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Page 158: ......
Page 178: ...Appendix B RS 232 Commands B 20 µ ...
Page 184: ......
Page 186: ...C 2 ...
Page 187: ...C 3 ...
Page 188: ...C 4 ...
Page 189: ...C 5 ...
Page 190: ...C 6 ...
Page 191: ...C 7 ...
Page 192: ...C 8 ...
Page 193: ...C 9 ...
Page 194: ...C 10 ...
Page 195: ...C 11 ...
Page 196: ...C 12 ...
Page 197: ...C 13 ...
Page 198: ...C 14 ...
Page 199: ...C 15 ...
Page 200: ......
Page 201: ...APPENDIX D PUMP MAINTENANCE ...
Page 202: ......
Page 203: ......
Page 204: ......