Chapter 9 Optional Equipment
REMOTE ACTIVATION OF ZERO/SPAN AND SAMPLE VALVES
The rear panel I/O (DB25) connector, shown in Figure 9-3, enables the zero/span and
sample valves to be remotely controlled via contact closure. In addition, the connector
has several instrument status outputs. Option switch 1 must be on and option switch 2 off
in order to enable the remote I/O connector.
Pin Out
(1)
Ground
(13)
NC
(2)
NC
(14)
Ground
(3)
NC
(15)
NC
(4)
NC
(16)
NC
(5) INPUT - Zero Gas
(17) NC
(6) Ground
(18) INPUT - Span Gas
(7)
Relay
Common
(19)
Ground
(8) STATUS - Concentration Alarm
(20) Relay Common
(9) STATUS - Local or Remote Mode
(21) STATUS - Zero Mode
(10) STATUS - ppm or mg/m3 mode
(22) STATUS - Span Mode
(11) STATUS - General Alarm
(23) NC
(12) Relay Common
(24) STATUS - NC
(25) Relay Common
Figure
9-3.
Rear Panel I/O Connector
64P947-5
9-6
Summary of Contents for 48C
Page 4: ......
Page 12: ......
Page 14: ......
Page 18: ......
Page 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Page 112: ......
Page 116: ......
Page 120: ......
Page 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Page 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Page 138: ......
Page 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Page 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Page 158: ......
Page 178: ...Appendix B RS 232 Commands B 20 µ ...
Page 184: ......
Page 186: ...C 2 ...
Page 187: ...C 3 ...
Page 188: ...C 4 ...
Page 189: ...C 5 ...
Page 190: ...C 6 ...
Page 191: ...C 7 ...
Page 192: ...C 8 ...
Page 193: ...C 9 ...
Page 194: ...C 10 ...
Page 195: ...C 11 ...
Page 196: ...C 12 ...
Page 197: ...C 13 ...
Page 198: ...C 14 ...
Page 199: ...C 15 ...
Page 200: ......
Page 201: ...APPENDIX D PUMP MAINTENANCE ...
Page 202: ......
Page 203: ......
Page 204: ......