Chapter 3 Operation
3-73
Default Coefficients
The Default Coefficient Screen is used prior to a calibration
when there has been a bad calibration or new ranges have been set up.
To view the Default Coef Screen
From the Main Menu Select Service Mode
From the Service Mode Menu select Hi or Lo Multi Point Cal
From Hi or Lo Multi-Point Cal select Default Coef
Default Coef Screen
To use the Default Coef Screen
Press
ENTER
to set coefficients to their default values
The Analyzer will automatically reset the default values and save the default
parameters COEF 0= 1.000000, COEF 1= 0.000000, COEF 2= 0.000000.
COEF 0 1.000000
1 0.009000
2 0.000000
SET TO DEFAULT?
Summary of Contents for 48C
Page 4: ......
Page 12: ......
Page 14: ......
Page 18: ......
Page 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Page 112: ......
Page 116: ......
Page 120: ......
Page 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Page 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Page 138: ......
Page 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Page 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Page 158: ......
Page 178: ...Appendix B RS 232 Commands B 20 µ ...
Page 184: ......
Page 186: ...C 2 ...
Page 187: ...C 3 ...
Page 188: ...C 4 ...
Page 189: ...C 5 ...
Page 190: ...C 6 ...
Page 191: ...C 7 ...
Page 192: ...C 8 ...
Page 193: ...C 9 ...
Page 194: ...C 10 ...
Page 195: ...C 11 ...
Page 196: ...C 12 ...
Page 197: ...C 13 ...
Page 198: ...C 14 ...
Page 199: ...C 15 ...
Page 200: ......
Page 201: ...APPENDIX D PUMP MAINTENANCE ...
Page 202: ......
Page 203: ......
Page 204: ......