Appendix B RS-232 Commands
B-16
program no
This reports the analyzer's program number and the Link (communications) program
number. The example below shows that the installed processor program is 48 00000100
and the installed communication program (link) is 48L 00000100.
Send:
"program no"
Receive:
"program no processor 48 00000100 link 48L 00000100"
set default params
This command sets the instrument’s parameters to their default values. The example
below shows a typical response. NOTE: It is important to send the “set save params”
command after issuing this command.
Send:
"set default params"
Receive:
"set default params ok"
set save params
This command stores parameters in the EEPROM. It is important that each time
instrument parameters are changed, that this command be sent. If changes are not saved,
they will be lost in the event of a power failure. The example below saves the parameters
to EEPROM.
Send:
"set save params"
Receive:
"set save params ok"
screen
This reports the information currently being displayed on the instrument's front panel
display. The example below shows a typical response to this command.
Send:
"sc"
Receive:
"
"
"CO PPM
29.5"
"
"
"SAMPLE 14:25
REMOTE"
Summary of Contents for 48C
Page 4: ......
Page 12: ......
Page 14: ......
Page 18: ......
Page 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Page 112: ......
Page 116: ......
Page 120: ......
Page 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Page 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Page 138: ......
Page 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Page 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Page 158: ......
Page 178: ...Appendix B RS 232 Commands B 20 µ ...
Page 184: ......
Page 186: ...C 2 ...
Page 187: ...C 3 ...
Page 188: ...C 4 ...
Page 189: ...C 5 ...
Page 190: ...C 6 ...
Page 191: ...C 7 ...
Page 192: ...C 8 ...
Page 193: ...C 9 ...
Page 194: ...C 10 ...
Page 195: ...C 11 ...
Page 196: ...C 12 ...
Page 197: ...C 13 ...
Page 198: ...C 14 ...
Page 199: ...C 15 ...
Page 200: ......
Page 201: ...APPENDIX D PUMP MAINTENANCE ...
Page 202: ......
Page 203: ......
Page 204: ......