Chapter 4 Calibration
4-9
Where:
[CO]
OUT
=
diluted CO concentration at the output manifold, ppm
[CO]
STD
=
concentration of the undiluted CO standard, ppm
F
CO
=
flow rate of CO standard corrected to 25
°
C and 760 mm Hg, LPM
F
D
=
flow rate of dilution air corrected to 25
°
C and 750 mm Hg, LPM
Allow the Model 48C to sample this CO concentration standard until a stable response is
obtained.
2. From the Main Menu select “Service Mode”. From “Service Mode” select “Lo
Multi-Point Cal”. From “Lo Multi-Point Cal” Select “Cal Point 1”. Use
←
and
→
pushbuttons to move the cursor and the
↑
and
↓
pushbuttons to increment or
decrement the digit values until they match the concentration being introduced to
the analyzer. Press
ENTER
. The analyzer will perform a series of calculations and
save the new parameters.
3. Press
MENU
to back up a step in the Service Mode menu.
4. Repeat step 1 for a 50% concentration of the upper range.
5. Select “Cal Point 2”
6. Use
←
and
→
pushbuttons to move the cursor and the
↑
and
↓
pushbuttons to
increment or decrement the digit values until they match the concentration being
introduced to the analyzer. Press
ENTER
. The analyzer will perform a series of
calculations and save the new parameters.
7. Press
MENU
to back up a step in the Service Mode menu.
8. Repeat step 1 for a 20% concentration of the upper range.
9. Select “Cal Point3”
10. Use
←
and
→
pushbuttons to move the cursor and the
↑
and
↓
pushbuttons to
increment or decrement the digit values until they match the concentration being
introduced to the analyzer. Press
ENTER
. The analyzer will perform a series of
calculations and save the new parameters.
11. Press
MENU
to back up a step in the Service Mode menu.
12. Select “Coefficients”.
13. Press
ENTER
. The analyzer will automatically calculate the new coefficients and
save the new parameters.
14. Repeat steps 1 through 13 for “Hi Multi-Point Cal”
Summary of Contents for 48C
Page 4: ......
Page 12: ......
Page 14: ......
Page 18: ......
Page 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Page 112: ......
Page 116: ......
Page 120: ......
Page 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Page 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Page 138: ......
Page 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Page 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Page 158: ......
Page 178: ...Appendix B RS 232 Commands B 20 µ ...
Page 184: ......
Page 186: ...C 2 ...
Page 187: ...C 3 ...
Page 188: ...C 4 ...
Page 189: ...C 5 ...
Page 190: ...C 6 ...
Page 191: ...C 7 ...
Page 192: ...C 8 ...
Page 193: ...C 9 ...
Page 194: ...C 10 ...
Page 195: ...C 11 ...
Page 196: ...C 12 ...
Page 197: ...C 13 ...
Page 198: ...C 14 ...
Page 199: ...C 15 ...
Page 200: ......
Page 201: ...APPENDIX D PUMP MAINTENANCE ...
Page 202: ......
Page 203: ......
Page 204: ......