Chapter 5 Preventive Maintenance
External leaks
To test for the presence of leaks around the fittings, disconnect the sample input line and
plug the
SAMPLE
fitting. The flow as read in the Flow screen (Diagnostics menu) should
slowly decrease to zero. The pressure as read in the Pressure screen (Diagnostics menu)
should drop to below 250 mm Hg. If the pump diaphragm is in good condition and the
capillary not blocked, it should take less than one minute from the time the inlet is
plugged to the time the reading below 250 mm Hg is obtained.
Leaks Across the Optional Zero/Span and Sample Solenoid Valves
In order to check for leaks across the optional valves, plug the
SPAN
inlet line, press the
RUN
pushbutton and follow the procedure above. If the pressure drops below 250 mm
Hg the valve associated with the span line is okay. Repeat for the valve associated with
the zero line by plugging the zero inlet, press the
RUN
pushbutton and follow the
procedure above. If the pressure drops below 250 mm Hg. the valve associated with the
zero line is okay.
5-3
Summary of Contents for 48C
Page 4: ......
Page 12: ......
Page 14: ......
Page 18: ......
Page 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Page 112: ......
Page 116: ......
Page 120: ......
Page 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Page 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Page 138: ......
Page 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Page 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Page 158: ......
Page 178: ...Appendix B RS 232 Commands B 20 µ ...
Page 184: ......
Page 186: ...C 2 ...
Page 187: ...C 3 ...
Page 188: ...C 4 ...
Page 189: ...C 5 ...
Page 190: ...C 6 ...
Page 191: ...C 7 ...
Page 192: ...C 8 ...
Page 193: ...C 9 ...
Page 194: ...C 10 ...
Page 195: ...C 11 ...
Page 196: ...C 12 ...
Page 197: ...C 13 ...
Page 198: ...C 14 ...
Page 199: ...C 15 ...
Page 200: ......
Page 201: ...APPENDIX D PUMP MAINTENANCE ...
Page 202: ......
Page 203: ......
Page 204: ......