Appendix B RS-232 Commands
B-4
co bkg
This command reports the current CO background. The example below reports that the
CO background is 1.4 ppm.
Send:
"co bkg"
Receive:
"co bkg 1.400 ppm"
set co bkg
dd.ddd
This command is used to set the CO background to a user-defined value. The example
below sets the CO background to 1.4 ppb.
Send:
"set co bkg 1.400"
Receive:
"set co bkg 1.400 ok"
set cal co bkg
This command performs the automatic calibration of the CO background. The example
below shows a successful automatic calibration.
Send:
"set cal co bkg"
Receive:
"set cal co bkg ok"
co coef
low co coef
The “co coef” command reports the co coefficient in single range mode and the high co
coefficient in dual range or auto range mode when using a zero/span calibration. The
example below reports a co coefficient of 1.005.
Send
“co
coef”
Receive
“co coef 1.005”
Summary of Contents for 48C
Page 4: ......
Page 12: ......
Page 14: ......
Page 18: ......
Page 27: ...Chapter 3 Operation 3 3 Figure 3 2 Flowchart of Menu Driven Software ...
Page 112: ......
Page 116: ......
Page 120: ......
Page 123: ...Chapter 7 Servicing 7 3 Figure 7 2 Model 48C Component Layout 48P737 ...
Page 125: ...Chapter 7 Servicing 7 5 Figure 7 3 Optical Bench Exploded View 42P761 ...
Page 138: ......
Page 146: ...Chapter 9 Optional Equipment 9 2 Figure 9 2 Model 48C Dimensional Outline D42P725 ...
Page 155: ...Chapter 9 Optional Equipment Figure 9 8 Purge Housing Assembly 9 11 ...
Page 158: ......
Page 178: ...Appendix B RS 232 Commands B 20 µ ...
Page 184: ......
Page 186: ...C 2 ...
Page 187: ...C 3 ...
Page 188: ...C 4 ...
Page 189: ...C 5 ...
Page 190: ...C 6 ...
Page 191: ...C 7 ...
Page 192: ...C 8 ...
Page 193: ...C 9 ...
Page 194: ...C 10 ...
Page 195: ...C 11 ...
Page 196: ...C 12 ...
Page 197: ...C 13 ...
Page 198: ...C 14 ...
Page 199: ...C 15 ...
Page 200: ......
Page 201: ...APPENDIX D PUMP MAINTENANCE ...
Page 202: ......
Page 203: ......
Page 204: ......