Preparing to Perform a Procedure
64
Spectra Optia Apheresis System Operator’s Manual
Selecting the Procedure
You must select the type of procedure you want to perform before you load the tubing set. Perform the following
steps to select a procedure:
1
Touch
Select Procedure
. The procedure selection screen appears.
2
Select the procedure you want to perform.
3
Touch
Confirm
. Once the system finishes loading the procedure software. a screen appears that displays the
filler and the tubing set required for the procedure.
Installing the Filler
Before you load the tubing set, verify that the correct filler is installed in the centrifuge. To remove and install a
filler, perform the following steps:
Removing the Filler
1
Open the centrifuge door.
2
Push the filler latch pin (Figure 5-1, number 1) toward the center of the centrifuge and raise the filler latch.
3
Push the filler locking pin (Figure 5-1, number 2) toward the center of the centrifuge. With the pin pushed
in, grasp the edge of the filler and lift up to raise the filler off the centrifuge.
Figure 5-1:
Location of the filler latch pin (1) and the filler locking pin (2)
2
1