Troubleshooting
222
Spectra Optia Apheresis System Operator’s Manual
Resuming a Procedure After a System Reset
The system resets itself under the following conditions:
•
Power to the system fails and is then restored.
•
You touch
Reset
, according to instructions on an alarm screen.
•
An alarm condition causes the system to reset itself.
•
You turn the system off and then on again.
After the reset, the system performs a series of tests to ensure that it is safe to resume the procedure. A screen
appears with instructions for how to proceed after the reset. Do the following:
1
Follow the instructions on the screen to resume the procedure at the point before the reset occurred. You may
also choose to discontinue the procedure.
• If the reset occurred during a run and you chose to continue the procedure, a screen appears asking you to
ensure that both saline lines are closed. Follow the instructions on the screen to confirm that the lines are
closed.
• If an alarm was active when the reset occurred, the action buttons that appear on the screen will be the
same as the buttons that were available on the alarm screen. If the alarm condition still exists after you
complete the instructions on the screen, the alarm will recur.
2
Verify that the system restored the patient and procedure data you entered before the reset occurred. The
system restores the data only if the tubing set cassette was in the lowered position when the reset occurred.
• If the reset occurred before you started a custom prime, the system does not restore the custom prime data.
Go to the screen that instructs you to enter the custom prime data and re-enter the data.
3
Verify that you completed the task that appeared on the screen before the reset occurred. Depending on
when the reset occurred, you may need to repeat a sequence of steps.
• If the reset occurred during administration of a fluid bolus, the system discontinued the bolus. Go to the
exchange status screen to view the volume that the system administered.
Note:
If power to the device has been off for more than 8 hours, the system might not allow you to resume the
procedure.