Spectra Optia Apheresis System Operator’s Manual
135
Mononuclear Cell (MNC) Collection Procedures
10
Inlet flow rate (mL/min)
5 to 120
Volume (mL)
100 to 400
(Volume should be at least 200 mL
for the patient to receive the full
benefit.)
Run values
AC infusion rate (mL/min/L TBV)
0.2 to 2.5.
(Caution status applies if the entry
exceeds 1.2.)
Inlet:AC ratio (__:1)
2 to 50
(Warning appears if the entered ratio
is not within the recommended range
of 6:1 to 15:1.)
Whole blood processed (mL)
Patient’s TBV multiplied by 1 to 10
Run time (min)
40 to 480
TBV processed
1 to 10
Inlet flow rate (mL/min)
10 to 125
Plasma volume: Target (mL):
Depends on the volume of plasma
collected into the collection bag.
(Total volume of plasma collected
cannot exceed 100% of the patient’s
total plasma volume, and cannot
cause the target or actual fluid
balance to be less than 75% of the
patient’s TBV.)
Collect flow rate (mL/min)
0.5 to 10
(Flow rate can be changed only when
the system is operating in
Semi-Automatic mode.)
Collect volume: Target (mL)
• Volume (mL)
• Number of collection phases (
)
Depends on the total volume of
plasma that was collected.
(Sum of the total plasma volume and
the collect volume cannot exceed
100% of the patient’s total plasma
volume, and the collect volume must
be less than or equal to the volume
of 50 collection phases.
Plasma volume in collection bag: Target (mL)
Depends on the volume of plasma to
be collected into the plasma bag.
(Total volume of plasma collected
cannot exceed 100% of the patient’s
total plasma volume, and cannot
cause the target or actual fluid
balance to be less than 75% of the
patient’s TBV.)
Table 10-2:
Data entry ranges for MNC collection procedure parameters (continued)
Screen
Parameter
Range