Spectra Optia System Specifications
250
Spectra Optia Apheresis System Operator’s Manual
Extracorporeal Volume (ECV)
Tubing Set and Accessory Set Components
Note:
Under normal operating conditions, the ECV does not exceed the typical ECV shown in Table 18-25. Under
certain infrequent alarm conditions, the ECV may momentarily increase to the maximum ECV unless the patient’s TBV is
less than 750 mL. If the patient’s TBV is less than 750 mL, the ECV does not exceed the typical ECV under any
condition.
Table 18-25:
ECV by procedure
Procedure
Tubing Set
Filler
Typical ECV (mL)
Maximum ECV
(mL)
TPE:
• Dual-needle access
• Single-needle access
• Exchange
• Exchange
• Standard
• Standard
• 141
• 185
• 185
• 185
RBCX
Exchange
Standard
141
185
TPE-SPD
Exchange
Standard
141*
185*
MNC collection
Collection
Standard
147
191
CMNC
IDL
IDL
253
297
PMN collection
IDL
IDL
253
297
WBCD
IDL
IDL
253
297
PLTD
IDL
IDL
253
297
*ECV shown does not include the volume of the secondary plasma device used during the procedure.
Note:
Single-use tubing sets for the Spectra Optia system are considered applied parts because they make contact
with the patient. The tubing sets comply with the Class II Type BF electrical safety requirements of IEC 60601-1.
Table 18-26:
Tubing set and accessory set components
Characteristics
Performance
Conditions
Remove bag volume
6 L
Bag is a component of the Exchange Set.
Collection bag volume
940 mL
Bag is a component of the Collection Set and
the IDL Set.
Plasma bag volume
• Liquid: 1,000 mL
• Frozen: 600 mL
• Bag is a component of the Collection Set, the
IDL Set, and the BMP Accessory Set.
• Volume shown is the maximum target
volume.
BMP bag volume
3 L
Bag is a component of the BMP Accessory Set.