Spectra Optia Apheresis System Operator’s Manual
203
Bone Marrow Processing (BMP) Procedures
14
Return to BMP bag and collect into bag buttons
The position of the collect valve determines whether the system is directing the flow of cells in the collect line back
to the BMP bag or into the collection bag. The system indicates the current position of the collect valve by showing
the corresponding button on this screen in a touched (lowered) position. You may manually change the position of
the collect valve at any time during the run.
•
To direct the flow of cells in the collect line back to the BMP bag, touch the return to BMP bag button and
then touch
Apply
. The collect valve moves into the return position, and the system begins pumping the cells
into the reservoir for return to the BMP bag.
•
To direct the flow of cells in the collect line into the collection bag, touch the collect into bag button and then
touch
Apply
. The collect valve moves into the collect position and the system begins pumping the cells into
the collection bag.
Note:
If you touch the return to BMP bag button, the system operates in Caution status because cells are not
being collected. The collect valve stays in the return position until you touch the collect into bag button.
SpectraOptiaOpsManual.book Page 203 Thursday, May 17, 2018 11:24 AM