Spectra Optia Apheresis System Operator’s Manual
257
Spectra Optia System Specifications
Table 18-32:
Voltage variations 60601-1-2 (3rd Ed.)
Immunity Test
Test Level
IEC 60601-1-2
Compliance Levels
Electromagnetic Environment
100 V
230 V
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines
IEC 61000-4-11
< 5% UT
(> 95% dip in UT) for
0.5 cycles
< 40% UT
(> 60% dip in UT) for
5 cycles
< 70% UT
(> 30% dip in UT) for
25 cycles
< 5% UT
(> 95% dip in UT) for
5 seconds
< 5% UT
(> 95% dip in UT) for
0.5 cycles
< 70% UT
(> 30% dip in UT) for
2 cycles
< 85% UT
(> 15% dip in UT) for
25 cycles
< 5% UT
(> 95% dip in UT) for
5 seconds
< 5% UT
(> 95% dip in UT) for
0.5 cycles
< 40% UT
(> 60% dip in UT) for
5 cycles
< 70% UT
(> 30% dip in UT) for
25 cycles
< 5% UT
(> 95% dip in UT) for
5 seconds
• Quality of the main power supply
should be that of a typical blood
center or a hospital environment.
• Interruptions of the main power
supply at voltages of 100 or less
will cause the system to restart
and the procedure to resume as
permitted by IEC 61000-4-11.
• If power interruptions are
frequent, the use of a suitably
rated uninterruptible power supply
is recommended.
Note: UT is the voltage of the main alternating-current power supply before use of the test level.