Ending the Run and Completing the Procedure
208
Spectra Optia Apheresis System Operator’s Manual
Ending the Run Before a Run Target is Attained
If you decide to end the run before a run target is attained, perform the following steps:
1
Touch the
End Run
menu button.
2
Choose one of the following options:
• To perform rinseback, touch the
Rinseback
tab.
• To skip rinseback, touch the
Disconnect
tab.
3
Follow the instructions on the screens to complete the procedure and disconnect the patient.
Ending the Run After a Run Target is Attained
The run targets screen appears and the system sounds a tone when one or more run targets have been attained. The
screen displays the target and current values of the run targets. The volume of AC in the product bags is also
displayed on the screens of the collection procedures. A yellow frame around a current value indicates that the run
target was attained. Review the run targets and decide to end or extend the run.
Ending the Run
Perform the following steps to end the run:
1
Touch
Rinseback
and follow the instructions on the screen to confirm your selection.
• If you are not performing rinseback, touch
Disconnect
and follow the instructions on the screen to
confirm your selection.
• If you decide not to perform rinseback, follow the instructions to skip rinseback under “Ending the Run
Before a Run Target is Attained” above.
2
Follow the instructions on the screen to clamp the inlet line.
3
Touch
Continue
. The system tests the pressure in the inlet line.
4
Follow the instructions on the screen to open the inlet saline line and to clamp and then seal the lines to the
product bags. See “Sealing the lines using the Seal Safe System” on page 210 for instructions on using the
Seal Safe system.
Warning:
Terumo BCT does not recommend performing rinseback during RBCX procedures. The data that the
system uses to predict the run targets does not include rinseback volume. If rinseback is performed, the run
targets may not be accurate.
Caution:
When performing a collection procedure using the Collection Set or the IDL Set, seal the collect line above the T-
shaped connector to the plasma line as shown in the illustration below. If you seal the collect line below the connector, the cells
in the collection bag could flow back into the tubing set when the system raises the cassette.
SpectraOptiaOpsManual.book Page 208 Thursday, May 17, 2018 11:24 AM