Spectra Optia Apheresis System Operator’s Manual
181
White Blood Cell Depletion (WBCD) and Platelet Depletion (PLTD) Procedures
13
Entering and Changing Data During the Run
The Spectra Optia system requires that you enter specific data, such as patient data, to perform a WBCD or a
PLTD procedure. The system also allows you to change numerical values for certain operating parameters during
the run. Table 13-2 shows data entry ranges for the operating parameters according to the screen on which each
data entry button appears. To change or enter data, touch the button for the parameter that you want to change. A
data entry pad appears to enable you to enter a number. The data entry pad displays either the established range for
the parameter or the calculated range, if the range changes based on related data. Also, the range for Caution status
is displayed in yellow, if the data you enter could cause the system to operate in Caution status.
Table 13-2:
Data entry ranges for WBCD and PLTD procedures
Screen
Parameter
Range
Patient data
Sex
• Male
• Female
Height:
• ft
• in
• cm
• 1 to 8
• 0 to 11
• 30 to 244
Weight:
• lb
• kg
• 5 to 500
• 2 to 227
TBV (mL)
300 to 15000
Hct (%)
10 to 70
WBCD procedures: WBC (E3/µL)
1 to 1000
PLTD procedures: Plt (E3/µL)
10 to 7000
Options
Rinseback
• Yes
• No
• Yes if a custom prime is not
selected
• No if a custom prime is selected
Custom prime
• Yes
• No
Default: No
HES
Selection determines the default packing factor.
• Yes: Default packing factor is 1.6
• No: Default packing factor is 4.5
• Yes
• No
Saline rinse
• Yes
• No
Default: No
Blood warmer: Return line
• Yes
• No
Blood warmer: Tubing set volume (mL)
1 to 100