22
5.4 Time (Optional)
First two digits
: hour
Last two digits
: minute
6. Power Management Settings:
6.1 Empty Weighing Plate Indicator (On)
If it is ON : There will be a flashing dot on the display instead of weight value if scale is
powered by battery and left empty for 30 seconds. ( “------” will be displayed on LCD models)
If it is OFF : Does not function.
6.2 Power Management (On)
Manages devices power management modes.
If it is ON : Power saving mode is active for all modes.
If it is OFF : Power saving mode is not active. If OFF “------” mode is also canceled.
7. Filters:
7.1 Median Filter (1)
If the scale is approved model, it cannot be changed. Some models may not have this menu.
Minimum value is 1, maximum value is 9. If a value is entered other than those, filter will use
its default value.
7.2 Average Length (50)
If the scale is approved model, it cannot be changed. Some models may not have this menu.
Minimum value is 10, maximum value is 50. If a value is entered other than those, filter will
use its default value.
7.3 Minimum Average Length (1)
If the scale is approved model, it cannot be changed. Some models may not have this menu.
Minimum value is 1, maximum value is 25. If a value is entered other than those, filter will
use its default value.
7.4 Filter Break Count Value
If the scale is approved model, it cannot be changed. Some models may not have this menu.
Minimum value is 70, maximum value is 2800. If a value is entered other than those, filter
will use its default value.
7.5 Automatic filter mode (wind filter) (Off)
If this mode is activated, 7.1 and 7.3 filter settings will be ignored. Device sets filter settings
automatically based on operating conditions.
8. Calibration:
8.1 Calibration Coefficient
It can be changed only if the calibration key is ON.
Summary of Contents for EX10200006
Page 1: ...USERS GUIDE for WEIGHING PRICE COMPUTING COUNTING SCALES ...
Page 3: ...2 PRICE COMPUTING SCALES ...
Page 15: ...14 WEIGHING SCALES Except EKO ...
Page 26: ...25 WEIGHING SCALES EKO ...
Page 33: ...32 COUNTING SCALES ...
Page 129: ...ACHTUNG Nach Verlassen des Setup Menüs schalten Sie die Waage aus und wieder ein ...
Page 135: ...UWAGA Po wyjściu z menu ustawień należy wyłączyć wagę i włączyć ją ponownie ...