Acquisition
Note:
You will undo all new measurements or math function definitions entered since the Auto
Setup when you Undo the Auto Setup. Perform this work when the instrument is not in the
Auto Setup mode if you wish for it to persist.
Restore Default Setup
Restore the factory default state by pressing the front panel
Default Setup
button. You can also
restore default settings by choosing
File > Recall Setup > Recall Default
.
Setting
Default
Channel/Vertical
C1 on at 50 mV/div Scale, 0 V Offset
Timebase
HD Mode on, Real Time Sampling at 50 ns/div, 0 Delay, 20 kS at 40
GS/s, 2.5 MS Memory
Trigger
C1 with an Auto Positive Edge, DC Coupling, 0 V Level
Display
Auto Grid
Cursors
Off
Measurements
Cleared
Math
Cleared
Viewing Status
All instrument settings can be viewed through the various Status dialogs. These show all existing
acquisition, trigger, channel, math function, measurement and parameter configurations, as well as
which are currently active.
Access the Status dialogs by choosing the Status option from the Vertical, Timebase, Math, or
Analysis menus (e.g., Channel Status, Acquisition Status).