HDO9000 High Definition Oscilloscopes Operator's Manual
Auxiliary Output
Use the Aux Output dialog to configure the output of the
Aux Out
port.
Auxiliary Output
Trigger Enabled
sends a pulse when the trigger is ready (Ready indicator lit), but not necessarily
triggered. It can be used as a gating function to trigger another instrument. Enter the desired pulse
Amplitude
.
Trigger Out
sends a pulse upon a trigger (TRIG'D Indicator lit). Enter the desired pulse
Amplitude
.
Pass/Fail
generates a pulse when Pass/Fail testing is active and conditions are met. Enter the
desired
Pulse Duration
.
Fast Edge
provides a fast edge signal that can be used to deskew multiple channels.
Off
disables auxiliary input/output.
Calibration Output
Choose the signal to be output from the Cal Out hook on the front of the instrument:
l
Square
wave signal. Enter the wave
Frequency
and
Amplitude into 1 MΩ
, or choose to
Set to 1
kHz, 1 V Square Wave
.
l
Reference
DC Level
. Enter an
Amplitude into 1 MΩ
.
Off
disables calibration output.