Operation Model
7600
Teledyne Analytical Instruments
64
5.
From the measurement mode, press the MODE key to
display the user mode.
6.
Point the cursor to “Setting of Auto Calibration” by
pressing the UP or DOWN key and press the ENT key.
7.
In the “Setting of Auto Calibration” item selection screen
that appears, point the cursor to “Auto Calibration Run”
by pressing the
UP or DOWN
keys. Press the ENT
key.
8.
“Stop” is highlighted, displaying a message to confirm
the stop of auto calibration. Press the ENT key to stop the
auto calibration, or press the ESC key to cancel the abort.
Note: During auto calibration, the keys are inoperable other than
key lock ON/OFF and “Stop Auto Calibration.”
When the key lock is set to ON, even the “Auto Calibration
Stop” is locked out. To stop “Auto Calibration” forcedly, set
the key lock to OFF and then execute “Auto Calibration
Stop” as detailed above.
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...