3.5.5.2.2
F
ORCIBLY
E
ND AN
A
UTO
Z
ERO
C
ALIBRATION
To abort an auto zero calibration event:
1.
From the measurement mode, press the MODE key to
display the user mode.
2.
Point the cursor to “Setting of Auto Calibration” by
pressing the UP or DOWN key and press the ENT key.
3.
In the “Setting of Auto Calibration” item selection screen
that appears, point the cursor to “Auto Calibration Run”
by pressing the
UP or DOWN
keys. Press the ENT
key.
4.
“Stop” is highlighted, displaying a message to confirm
the stop of auto calibration. Press the ENT key to stop the
auto calibration, or press the ESC key to cancel the abort.