Infrared Gas Analyzer
Operation
Teledyne
Analytical
Instruments
39
The numbered call outs in Figure 3-4 refer to:
No.
Name Function
(1) Component
display
Displays component of instantaneous
value, corrected instantaneous value,
corrected average value, etc.
(2) Concentration
display
Displays measured value of concentration.
(3) Range display Displays range values.
(4) Unit display
Displays unit with ppm and
vol %.
(5) Peak
alarm
component
Displays peak alarm component.
(6) Peak
alarm
concentration
Displays peak alarm concentration display.
(Upper limit value)
(7) Peak
alarm
times
Displays the alarm times exceeding the
peak value.
(8) Peak
alarm
unit display
Displays units of peak alarm with times/H.
Instantaneous value and concentration value:
The concentration display of Ch (component) where sampling
components such as “CO
2
”, “CO” or “O
2
are displayed in the component
display, indicates current concentration values of the measured
components contained in gas that is now under measurement.
O
2
correction concentration values:
Ch components where “cv**” is displayed as “cv CO” in the
component display are calculated values. They are obtained from the
following equation by setting sampling components, O
2
instantaneous/
concentration values and O
2
correction reference value (see item 6.8).
Cs
Os
21
On
-
21
output
Correction
×
⎟
⎠
⎞
⎜
⎝
⎛
−
=
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...