Infrared Gas Analyzer
Installation
Teledyne
Analytical
Instruments
29
•
The dew point of the sample gas must be lower th
an
the
ambient temperature to avoid condensation in the analyzer. If
vapor is contained in the sampling gas, the dew point should
be lowered to 0°C by using a dehumidifier.
•
If SO
2
mist is contained in the sample gas, use a mist filter or
cooler to remove the SO
2
mist. Other mists should be
removed by using an appropriate mist filter or cooler.
•
Corrosive gases such as Cl
2
, F
2
an
d
HCl, if contained in the
sample gas in considerable amounts, will sho
rt
en the life of
the instrument.
•
Sample gas temperature should be within 0 to 50°C. Hot gas
should not be fed directly into the instrument.
2.5.4 Flowrate
Use a flowrate of 0.5L/min ±0.2L/min. The sample system should
be designed to avoid any flow fluctuation du
ri
ng measurement.
Use a flowmeter to observe the flow reading as shown in the
external piping diagrams of Figures 2.5, 2.6 and 2.7.
2.5.5 Preparation of Calibration Gas
Routine calibration using calibration gases is required for
optimizing the performance of this instrument. Once a week is a
suggested calibration frequency. Table 2-2 indicates the zero and span
gas required for calibration.
Table 2-2: Zero and Span Gas
Analyzer without
O
2
Analyzer with built-in
O
2
Analyzer with external
zirconia O
2
sensor
Zero gas
N
2
gas
N
2
gas
Dry air
Span gas
other than for
O
2
t
Gas concentration
of 90% or more of
full scale
Gas concentration of
90% or more of full
scale
Gas concentration of 90%
or more of full scale
Span gas for
O
2
analysis
NA
Gas concentration of
90% or more of full
scale
1 to 2% O
2
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...