Infrared Gas Analyzer
Appendix
Teledyne
Analytical
Instruments
115
A.3 Optional Functions
O
2
correction:
Correction of measured NO, SO
2
and CO
gas concentrations into values at reference
O
2
concentration
Correction
formula:
Cs
Os
-
21
On
-
21
C
×
=
C: Sample gas concentration after O
2
correction
Cs: Measured concentration of sample gas
Os: Measured O
2
concentration (Limit setting:
1 to 20% O
2
)
On: Reference O
2
concentration (value
changeable by setting. 0 to 19% O
2
)
Average value after O
2
correction and O
2
average value calculation:
The result of O
2
correction or instanta-
neous O
2
value can be output as an av-
erage value in the preset period of time.
The moving average method is used for
averaging in which sampling is carried out
at intervals of 30 seconds.
Output is updated every 30 seconds. It is
the average value in the determined period
of time just before the latest update.
Averaging time is settable within 1 to 59
minutes (in increments of 1 minute) or 1
to 4 hours (in increments of 1 hour).
Average value resetting:
The above-mentioned output of average
value is started from the initial state by
opening the average value resetting input
terminals after short-circuiting for 1.5 sec-
onds or longer.
Output is reset by short-circuiting and re-
started by opening
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...