![Teledyne 7600 Operating Instructions Manual Download Page 40](http://html.mh-extra.com/html/teledyne/7600/7600_operating-instructions-manual_1080467040.webp)
Operation Model
7600
Teledyne Analytical Instruments
40
Where:
On: The value of the O
2
correction reference value (Value set by
application)
Os: Oxygen concentration (%)
Cs: Concentration of relevant measured component. Note that Os
does not exceed the O
2
limit value set in “Other Parameter”
See Section 3.6
Maintenance Mode.
The corrected sampling components are NO
X
, SO
2
and CO only.
O
2
correction concentration average value:
In the Ch (component) and O
2
average value where “CV/AV **” is
displayed as “ AV CO” in the component display, a value obtained by
averaging O
2
correction concentration value or O
2
average value in a
fixed time is output every 30 seconds.
Averaging time can be changed between 1 minute and 59 minutes
or 1 hour and 4 hours according to the average time settings (See
Section 3.5.7
Parameter Setting
).
The set time is displayed as “1h” , for instance, in the range display.
* The measurement ranges of O
2
correction concentration value and O
2
correction concentration average value are the same as that of the
measuring components. Also, the measurement range of O
2
average
value is the same as that of O
2
.
3.3.2 Setting/Selection Screen
The setting/selection screen is configured as shown in Figure 3-5:
•
In the status display area, the current status is displayed.
•
In the message display area, messages associated with
operation are displayed.
•
In the setting item and selection item display area, items or
values to be set are displayed, as required. To work in a
particular area, move the cursor to any item by using the UP,
DOWN and LEFT/RIGHT keys.
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...