Infrared Gas Analyzer
Operation
Teledyne
Analytical
Instruments
47
4.
In the “Calibration Parameters” screen that appears, point
the cursor to “Calibration Value” by pressing the UP or
DOWN key. Press the ENT key.
To change the concentration of a calibration gas:
1.
Point the cursor to the channel you want to set by using
the
UP or DOWN
key. Press the ENT
key.
2.
Select the concentration item you want to set by pressing
the UP or DOWN or SIDE key. Press the ENT key, and
the selected value is highlighted.
3.
Enter the calibration gas concentration values (zero and
span). For any numerical value entry, press the UP or
DOWN key to increase or decrease a single digit. Select
the next digit by pressing the SIDE key.
4.
When the concentration is correct, save the entry by
pressing the ENT key. The saved value becomes valid
from the next calibration process.
Summary of Contents for 7600
Page 14: ...Model 7600 Teledyne Analytical Instruments xiv ...
Page 20: ...Introduction Model 7600 Teledyne Analytical Instruments 20 ...
Page 96: ...Maintenance Model 7600 Teledyne Analytical Instruments 96 Figure 4 3 Sample Cell ...
Page 108: ...Maintenance Model 7600 Teledyne Analytical Instruments 108 ...
Page 120: ...Appendix Model 7600 Teledyne Analytical Instruments 120 A 9 Description on Terminal Block ...
Page 121: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 121 ...
Page 122: ...Appendix Model 7600 Teledyne Analytical Instruments 122 ...
Page 123: ...Infrared Gas Analyzer Appendix Teledyne Analytical Instruments 123 ...
Page 124: ...Appendix Model 7600 Teledyne Analytical Instruments 124 ...
Page 126: ...Index Model 7600 Teledyne Analytical Instruments 126 ...